Thermo Scientific TSQ 9000 Hardware Manual page 270

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
242
TSQ 9000 Hardware Manual
10. Set the manifold door right side up on a clean work surface and remove the two screws
holding the black pillow block in place.
Figure 248. Removing the Clamp
Black Pillow Block
Screws
11. Remove the three screws holding the metal valve body to the front manifold plate.
Figure 249. Removing the Vacuum Interlock Handle
Screws
Foreline Hose
Front
Thermo Scientific