Components Of The Source Exchange Tool - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
2. Get the source exchange tool that shipped with your instrument. It is used to remove and
insert the ion source cartridge through the vacuum interlock. The tool has a large black
handle on one end and a stainless steel barrel at the other. On the black handle is a
diagram describing how to insert and remove the ion source cartridge from the
instrument using the tool. The stainless steel barrel is cut with a track and the black
handle moves up and over the track, depending on whether you are removing or installing
an ion source cartridge.
Figure 32. Components of the Source Exchange Tool
3. Grasp the handle of the source exchange tool in one hand and use your other hand to pull
the metal barrel out and away from the handle so that tool is fully extended.
4. Attach the source exchange tool to the front of the instrument.
Handle
Usage Diagram
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Left Track
of Barrel
Right Track
of Barrel
TSQ 9000 Hardware Manual
Alignment
Groove
45

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents