11. Twist the vent valve knob to remove it and the vent valve o-ring.
Figure 279. Removing the Vent Valve Knob and O-Ring
Vent Valve
O-ring
Vent
Valve
Knob
12. Flip the manifold door over and remove the manifold o-ring.
Figure 280. Removing the Manifold O-Ring
Manifold
O-ring
13. Twist and remove the source magnet retainers that hold the magnet yoke to the back of
the manifold door. The magnet yoke will pull free of the door, releasing the alignment
collar.
Figure 281. Removing the Magnet Yoke and Source Magnet Retainers
Magnet Yoke
Source Magnet
Retainers
3
Advanced Troubleshooting
Replacing Components of the Manifold
TSQ 9000 Hardware Manual
263