Advanced Troubleshooting

Replacing Components of the Source Exchange Tool

Replacing Components of the Source Exchange Tool
Figure 234. Replaceable Components of the Source Exchange Tool
234
TSQ 9000 Hardware Manual
10. For optimal performance, wait at least 30 minutes for the ion source to heat up to the
same temperature as the inside of the instrument. Otherwise, the masses or intensities
may be different
.
IMPORTANT When inserting a cold ion source cartridge such as after cleaning, the ion
source and lens stack will expand as the source cartridge heats, often pushing the ion
volume and lenses away from the rear of the instrument where they are firmly held by the
RF Lens spring contacts. Inserting the source with the gap shown in the tip above can help
prevent contact issues.
If the components of your source exchange tool get worn or damaged, you can use the
illustration below to visually locate a component and then follow the process to replace it.
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
Seal
Clip
Bushing