Attaching The Ci Reagent Gas Cable To Distribution Board - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
21. Attach the CI reagent gas controller cable to the distribution board.
Figure 465. Attaching the CI Reagent Gas Cable to Distribution Board
22. Reinstall the power module. Push down on the module and make sure it is flush against
the top of the chassis and the screw holes line up.
23. Reattach the two screws holding the power module to the instrument.
24. Reattach the controller interface board, the PC communication board, and the support
bracket.
25. Connect the Ethernet cable to the PC communication board.
26. Secure the left hand sub panel to the chassis with the three screws.
27. Plug in the power connections.
28. Plumb the reagent gas from your gas supply to either port on your CI gas controller.
WARNING FIRE HAZARD: Some CI gases, such as methane and isobutane, are
flammable. Make sure these gases are properly exhausted and all gas fittings on the system
are leak-free. Consult your local Environmental and Safety Regulations for information
about how to properly exhaust fumes from your laboratory.
4
Upgrade Equipment
CI Reagent Gas Flow Module
CI Reagent Gas
Flow Module
Connector
TSQ 9000 Hardware Manual
433

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents