Left Hand Front Panel And Left Hand Sub Panel; Locating The Left Hand Front Panel Screws - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Left Hand Front Panel and Left Hand Sub Panel

Thermo Scientific
To replace the left hand front panel and the left hand sub panel of the TSQ 9000
instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127. You do not have to remove the ion source cartridge or column or vent the
system. If you are using hydrogen as a carrier gas, power off the GC.
2. Remove the four screws on the left hand front panel. One is located on the bottom left of
the interior front panel of the instrument.
Figure 314. Locating the Left Hand Front Panel Screws
3. To replace the left hand sub panel, disconnect the inputs as shown in
3
Replacing the Covers of the TSQ 9000 Instrument
"Shutting Down the TSQ 9000 System"
Screws
TSQ 9000 Hardware Manual
Advanced Troubleshooting
Screw on Front
Figure
315.
293