Advanced Troubleshooting
Replacing Components of the Front Door

Front Door Hinges and Supports

300
TSQ 9000 Hardware Manual
To replace the hinges and supports that connect the front door to the chassis
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127. You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 320. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
3. Lift the left hand front panel off the instrument.
"Shutting Down the TSQ 9000 System"
Screws
Screw on Front
of Instrument
Thermo Scientific