Thermo Scientific TSQ 9000 Hardware Manual page 251

Table of Contents

Advertisement

Thermo Scientific
8. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door. See
Figure
217.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
Figure 217. Removing the Manifold Door Screws
9. Open the manifold door. You can now see the ion source block.
Figure 218. Locating the Ion Source Block
10. Insert the small source removal tool into the ion source block as shown in
3
Advanced Troubleshooting
Replacing Components of the Ion Source
Ion Source Block
Figure
219.
TSQ 9000 Hardware Manual
223