10. Reconnect the power cables on the left side of the instrument. Make sure you attach the
cables properly. The foreline pump connects to the System Mech Pump connector on the
left of the instrument.
11. Reattach the top cover panel and tighten the screw holding it in place. Reattach the
hydrogen safety screws if you are using hydrogen.
12. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing the Power Supplies
"Restarting the System"
on
TSQ 9000 Hardware Manual
page
145.