5. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately two inches from the top of the board directly under the white label.
Figure 268. Removing the Lens Driver Board
6. Use your fingers to carefully pull the bracket of the board away from the chassis and
toward the left side of the instrument, until the 20 pin connector clears the feedthrough.
7. Pinch together the two springs on the left side the vacuum interlock shield to free the
hinge from the vacuum interlock shield.
Figure 269. Freeing the Hinge from the Vacuum Interlock Shield
Springs
3
Advanced Troubleshooting
Replacing Components of the Manifold
Lens Driver