Advanced Troubleshooting
Replacing Components of the Ion Source
188
TSQ 9000 Hardware Manual
8. Remove repeller plate
Figure 171. Removing the Repeller Plate
9. Disconnect the repeller plate wire from the source interface board.
10. Connect the new repeller plate wire to the source interface board. Do not trap any
insulation in the source interface board connector.
11. Attach the new repeller plate.
12. Reattach the insulating spacers.
13. Reattach the source-repeller thumbscrews.
14. Reattach the manifold door and the four screws.
15. Reattach the top cover and tighten the screw.
16. Restart your system by following the instructions in
Repeller Plate
Repeller Plate Wire
"Restarting the System"
on
page
145.