Figure 174. Locating Wires on the Source Interface Board
5. Attach a new lens plate and springs to the lens heater block and reattach the screws.
6. Reattach the lens 1, lens 2, and lens 3 wires from the source interface board using a
flat-tipped screwdriver.
7. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 9000 Instrument"
instrument.
8. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Components of the Ion Source
on
page 142
to replace the tray in the TSQ
"Restarting the System"
TSQ 9000 Hardware Manual
"Replacing the
on
page
145.