Source Interface Board and Heat Shield

Thermo Scientific
To replace the source interface board and heat shield
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
127.
2. If you are only replacing the heat shield, you do not need to remove the analyzer tray.
Disconnect the filament board and wires from the source interface board and the left
screw to remove the heat shield.
Figure 377. Replacing the Heat Shield
Filament Board
Left
and Wires
Screw
Note Keep all manifold parts you remove covered on a clean work surface so they will not
accumulate dust.
3. Replace the heat shield.
4. Reconnect the filament board and wires to the source interface board.
5. If you are replacing the source interface board, remove the analyzer tray from the
instrument by following the instructions in
9000 Instrument"
on
Source Interface
Board
"Removing the Analyzer Tray from the TSQ
page
129.
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
"Shutting Down the TSQ 9000 System"
Heat Shield
TSQ 9000 Hardware Manual
351