Advanced Troubleshooting
Replacing Components of the Analyzer
132
TSQ 9000 Hardware Manual
Figure 94. Opening the Manifold Door
9. Open the manifold door and slide the filament board and wires off the pins on the source
interface board.
10. Remove the column and replace with a nut and no-hole ferrule.
11. Pull the GC away from the mass spectrometer.
12. Remove the instrument right-side panel to access the transfer line. See
Figure 95. Removing the Right Side Panel
13. Identify the set screw on the source heater block. This T10 screw holds the transfer line in
alignment with the ion volume heater block. See
Screws
Figure
96.
Figure
95.