7. Lower the handle and lift the vacuum interlock shield off the instrument. The notch at
the bottom of the track will go around the handle.
Figure 247. Detaching the Manifold Door
Screws
Lower the Vacuum
Interlock Handle
8. Use a T20 Torxhead screwdriver to remove the four screws found in the four corners of
the manifold door.
9. Use a flathead screwdriver to disconnect the foreline hose at the 90° hose connector.