Thermo Scientific TSQ 9000 Hardware Manual page 263

Table of Contents

Advertisement

Thermo Scientific
To replace the components of the source exchange tool:
1. Use a T20 Torxhead screwdriver to remove the three screws at the end of the black
handle.
Figure 235. Removing Source Exchange Tool Screws
Screws
2. Loosen the set screw at the other end of the black handle.
3. Slide the black handle toward the other end of the tool.
4. Use a T10 Torxhead screwdriver to remove the set screw holding the endcap onto the
interior shaft.
5. Remove the endcap.
Figure 236. Removing the Endcap
Endcap
6. Slide the handle off the rest of the tool. Set it aside.
7. Remove the interior shaft from the barrel.
Figure 237. Removing the Interior Shaft
Replacing Components of the Source Exchange Tool
Set Screw
Barrel
3
Advanced Troubleshooting
Set Screw
Interior Shaft
TSQ 9000 Hardware Manual
235