Advanced Troubleshooting
Replacing Components of the Ion Source
226
TSQ 9000 Hardware Manual
1. Attach the ion source cartridge to the end of the small source removal tool as shown in
Figure
222.
Note For correct insertion, the notch on the ion source cartridge must be facing up.
Figure 222. Attaching the Ion Source Cartridge to the Small Source Removal Tool
Notch on Ion Source Cartridge
2. With the notch on the ion source cartridge facing straight up, insert the ion source
cartridge into the instrument. See
Figure 223. Inserting the Ion Source Cartridge into the Instrument
3. Once the ion source cartridge is inserted into the ion source block, turn the small source
tool to the right until you feel the ion source cartridge engage.
4. Remove the small source tool from the instrument.
5. Be sure that the source interface board wires do not touch anything metal in the manifold
before closing the door.
Figure
223.