Replacing The Chassis Feet - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Replacing the Chassis Feet

Thermo Scientific
To replace the chassis feet on the bottom of the TSQ 9000 instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 9000 Instrument"
3. Follow the instructions in the section,
components required to safely replace the chassis feet.
4. Remove the small and large glass covers in the manifold.
5. Carefully tip the TSQ 9000 instrument and rest it on its left side.
6. Use a T20 Torxhead screwdriver to remove the screws holding any damaged feet to the
bottom of the instrument.
Figure 318. Removing the Feet on the Bottom of the TSQ 9000 Instrument
7. Attach the new chassis feet.
8. Set the instrument upright.
9. Replace the large and small glass manifold covers.
10. Follow the instructions in the section,
components you removed.
Replacing the Covers of the TSQ 9000 Instrument
"Shutting Down the TSQ 9000 System"
"Distribution Board"
on
Feet
"Distribution Board"
on
TSQ 9000 Hardware Manual
3
Advanced Troubleshooting
on
page
129.
page 337
to remove the
page 337
to replace the
297