Quad 1 And Quad 3 Wire Sets - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Analyzer

Quad 1 and Quad 3 Wire Sets

164
TSQ 9000 Hardware Manual
To replace the quad 1 or quad 3 wire sets
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 9000 Instrument"
3. For both quad assemblies, use a T10 Torxhead screwdriver to remove the four screws
around the rods. These screws hold the wires in place. If the screws won't come off the
rods, you must replace the whole quad assembly.
4. Use on the screws provided. These screws are made from a special material designed not
fuse with the quadrupole rods.
Figure 137. Replacing the Q1 Quad Assembly Wires
Screws
"Shutting Down the TSQ 9000 System"
on
page
129.
Screws
Thermo Scientific