Ion Guide Wire Set

Thermo Scientific

To replace the ion guide wires

Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
127.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 9000 Instrument"
3. Use a T10 Torxhead screwdriver to remove the screw connecting the ion guide wire to the
top of the ion guide.
Figure 131. Replacing the Ion Guide Wires
4. Use a wrench to remove the nut connecting the other ion guide wire to the side of the ion
guide.
5. Attach new ion guide wires and reattach the nuts and screw.
6. Follow the instructions in
page 142
to replace the tray into the instrument.
7. Restart your system by following the instructions in
"Shutting Down the TSQ 9000 System"
Ion Guide Wires
"Replacing the Analyzer Tray in the TSQ 9000 Instrument"
3
Advanced Troubleshooting
Replacing Components of the Analyzer
on
page
129.
"Restarting the System"
on
TSQ 9000 Hardware Manual
Nut
Screw
on
page
145.