Advanced Troubleshooting
Replacing Components of the Manifold
262
TSQ 9000 Hardware Manual
8. Set the manifold door right side up on a clean work surface and remove the two screws
holding the black clamp in place.
Figure 277. Removing the Clamp
Black Clamp
Screws
9. Remove the three screws holding the metal block to the front manifold plate.
Figure 278. Removing the Vacuum Interlock Handle
Screws
10. Set the vacuum interlock and all of its components aside.
Front
Note: Tube
not
attached
Thermo Scientific