3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
250
TSQ 9000 Hardware Manual
34. Reattach the manifold door and the four screws.
Figure 263. Reattaching the Manifold Door
Screws
35. Hold the spring and safety interlock in place as you attach the vacuum interlock shield
and screws.
36. Reattach the vacuum interlock knob.