Thermo Scientific TSQ 9000 Hardware Manual page 375

Table of Contents

Advertisement

Thermo Scientific
Figure 372. Disconnecting the Cable to the Bulkhead Fitting
Bulkhead
Fitting
28. Disconnect the collision gas module cable from the distribution board.
29. If you are using CI, follow the instructions in the
remove the CI reagent gas flow module.
30. Now that all of the components have been removed from the distribution board, you can
see that it lies flat on the bottom of the chassis.
Collision Gas Module
Tubing
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
CI Reagent Gas Flow Module
TSQ 9000 Hardware Manual
section to
347