Reattaching the Analyzer Tray Components

Thermo Scientific
14. Remove the Q1 exit, Q3 entrance, and Q1 exit lens wires from the analyzer tray. See
Figure
102.
15. Use a T20 Torxhead screwdriver to remove the two screws attaching the source block to
the analyzer tray.
IMPORTANT Be careful not to drop the MACOR ceramic plate. See
fragile.
Figure 106. Removing the Ion Source from the Analyzer Tray
Screws
16. Remove the source block from the analyzer tray and set it aside.
17. Replace the analyzer tray with a new one if needed.

To reattach the analyzer tray components to the tray

1. Reattach the source and lens heater block to the analyzer tray with the two screws.
2. Reattach the screw holding the lens heater block grounding strap to the analyzer tray.
3. Make sure the ion guide is aligned parallel to the flat on the endcap.
4. Tighten the T10 screws inside the endcap.
Note Make sure the Q1 entrance lens sits flush and centered in the Q1 entrance lens
endcap.
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Figure
106. It is
MACOR
Ceramic Plate
TSQ 9000 Hardware Manual
139