8. Flip the electron multiplier and plate over so that the plate is resting on the worktable.
Figure 303. Resting the Electron Multiplier Plate on the Worktable
Electron
Multiplier
Plate
9. Remove the two screws connecting the electron multiplier to the plate and attach the
electron multiplier to a new plate.
Figure 304. Replacing the Electron Multiplier Plate
Screws
10. Reattach the large feedthroughs, screws, and o-rings
11. Reattach the anode feedthrough, two screws, standoff. and o-ring
12. Move the wires into the large feedthroughs.
13. Put the electron multiplier plate back into place on the manifold and reattach the four
screws. Reattach the electrometer board and the four screws.
14. Reattach the electrometer shield and the two screws
3
Advanced Troubleshooting
Replacing Components of the Electron Multiplier
TSQ 9000 Hardware Manual
279