Thermo Scientific TSQ 9000 Hardware Manual page 358

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
330
TSQ 9000 Hardware Manual
21. Reinsert the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
Figure 354. Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board
22. Push the board down into the connector.
23. Look through the left side of the instrument and tighten the screw that holds the lens
driver board in place on the chassis.
24. Reattach the top cover panel and tighten the screw.
25. Restart your system by following the instructions in
"Restarting the System"
on
Thermo Scientific
page
145.