Manifold Door Hinge; To Replace The Manifold Door Hinge - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Manifold Door Hinge

Thermo Scientific

To replace the manifold door hinge:

Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127.
Note You do not have to remove the ion source cartridge or the column.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Remove the four screws on the left hand front panel and remove the panel. One is located
at the bottom left side of the front of the instrument.
Figure 266. Locating the Left Hand Front Panel Screws
3
Replacing Components of the Manifold
"Shutting Down the TSQ 9000 System"
Nuts
Screw Located on Front of
Instrument
TSQ 9000 Hardware Manual
Advanced Troubleshooting
253

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents