Advanced Troubleshooting
Replacing the Power Supplies
366
TSQ 9000 Hardware Manual
Figure 389. Replacing the Dynode/Multiplier Power Supply
12. Reattach the screw holding the dynode and multiplier power supply in place.
13. Connect the cables to the clips attached to the chassis
14. Connect the large feedthrough cables.
15. Replace the main power supply.
16. Reattach the top cover panel and tighten the screw.
17. Restart your system by following the instructions in
Connector
"Restarting the System"
on
Thermo Scientific
page
145.