Thermo Scientific TSQ 9000 Hardware Manual page 329

Table of Contents

Advertisement

Thermo Scientific
4. On the front interior panel, use a T20 Torxhead screwdriver to remove the hinge support
bracket screw.
Figure 321. Replacing the Hinges on the Front Door
5. Use a T10 Torxhead screwdriver to remove the all four screws connecting each hinge to
the front door and interior front panel.
6. Attach new hinges and supports, as well as the hinge support bracket screw.
Figure 322. Replacing the Hinge Supports
Hinge Support
Note Replace one hinge at a time to prevent the door from falling.
7. Reattach the left hand front and top cover panels and tighten the screws.
8. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Components of the Front Door
Hinge Screws
Hinge Support
Bracket Screw
"Restarting the System"
TSQ 9000 Hardware Manual
on
page
145.
301