Advanced Troubleshooting
Replacing Components of the Ion Source
230
TSQ 9000 Hardware Manual
Figure 226. Removing the Manifold Door Screws
8. Open the manifold door.
9. Let the system cool down for 30 minutes.
CAUTION - BURN HAZARD The ion source cartridge may be hot.
AVERTISSEMENT RISQUE DE BRÛLURE :. la cartouche de la source d'ions peut être
chaude.
10. Disconnect the filament wires from the source interface board.
Figure 227. Disconnect the Filament Wires
11. Grasp the Vespel cap and gently remove the Advanced EI ion source from the ion source
block. See
Figure
219.
Filament Wires