Lens Heater Block Grounding Strap

Thermo Scientific
To replace the lens heater block grounding strap:
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 9000 Instrument"
3. Use a T10 Torxhead screwdriver to disconnect the lens heater block grounding strap from
the lens heater block.
Figure 172. Disconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw
4. Attach a new lens heater block grounding strap and reattach the screws.
5. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 9000 Instrument"
instrument.
6. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Components of the Ion Source
"Shutting Down the TSQ 9000 System"
on
page
on
page 142
to replace the tray in the TSQ
"Restarting the System"
TSQ 9000 Hardware Manual
129.
"Replacing the
on
page
145.
189