Reinserting the Ion Source Cartridge

Thermo Scientific
Figure 54. Inserting the Locking Ring into the Sleeve
Larger Gaps on
Locking Ring
IMPORTANT When inserting a cold ion source cartridge such as after cleaning or when
switching between EI and CI modes, the ion source and lens stack will expand as the
source cartridge heats, often pushing the ion volume and lenses away from the rear of the
instrument where they are firmly held by the RF Lens spring contacts. To avoid
intermittent electrical contacts to the lenses, you should insert the ion source cartridge,
wait 30 minutes for it to get to temperature, then remove and reinsert it.
1. Hold the source holder and ion source cartridge in one hand and use the other hand to
hold the ion cartridge tool so that the barrel is facing toward the floor. The diagram on
the tool should be facing you.
IMPORTANT If you try to attach the ion source cartridge to the tool by inverting it,
the components will likely fall, so make sure you are working over a table. Otherwise,
you may have to stop and clean the components again.
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Hooks on Sleeve
TSQ 9000 Hardware Manual
61