Transfer Line - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Calibration Gas Components

Transfer Line

378
TSQ 9000 Hardware Manual
To replace the transfer line and ferrule
IMPORTANT Be sure to install the transfer line correctly or the ion source will not align
correctly.
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
127.
2. Open the front door of the GC and disconnect the column and transfer line nut.
3. Move the TSQ 9000 instrument away from the GC so you can access the transfer line on
the right side.
WARNING BURN HAZARD: The transfer line may be hot, so be careful.
4. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
Figure 402. Removing the Right Side Panel Screws
5. Slide the right side panel to the right of the instrument and lift it off.
6. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
"Shutting Down the TSQ 9000 System"
Right Side
Panel Screws
Thermo Scientific