Removing The Electrometer Shield; Removing The Electrometer Board - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
Figure 299. Removing the Electrometer Shield
Screws
5. Remove the four screws holding the electrometer board to the electron multiplier plate.
CAUTION - ELECTRICAL HAZARD Avoid dropping screws between chassis and side cover
or board. If you drop a screw, find it before powering the instrument on again.
Figure 300. Removing the Electrometer Board
Electron
Multiplier
Plate
Replacing Components of the Electron Multiplier
Screws
3
Advanced Troubleshooting
TSQ 9000 Hardware Manual
277