Advanced Troubleshooting
Replacing Components of the Electron Multiplier
282
TSQ 9000 Hardware Manual
Figure 306. Disconnecting the Electron Multiplier Plate
7. Flip the plate over and attach the wire to the large feedthrough.
8. Use a T20 Torxhead screwdriver to reattach the four screws at each corner of the electron
multiplier plate
9. Reconnect the large feedthrough cables.
10. Reattach the top cover panel and tighten the screw.
11. Restart your system by following the instructions in
Electron
Multiplier Plate
Screws
"Restarting the System"
on
Thermo Scientific
page
145.