PC Communication Board, Controller Interface Board, and Support Bracket
Thermo Scientific
To replace the PC communication board, controller interface board, or support

bracket

Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
127.
Note You do not have to remove the ion source cartridge or the column or vent the
instrument.
2. Slide the top cover panel off the instrument.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 355. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
CAUTION - INSTRUMENT DAMAGE Limit handling of the boards as much as possible.
Do not touch any board components. Hold the board by the sheet metal bracket only.
3
Replacing the Boards in the TSQ 9000 Instrument
"Shutting Down the TSQ 9000 System"
Screws
TSQ 9000 Hardware Manual
Advanced Troubleshooting
Screw on Front
of Instrument
331