Advanced Troubleshooting
Replacing Fans and Filters

Front Wire Mesh Filter

392
TSQ 9000 Hardware Manual

To replace the front wire mesh filter

Note If needed, you can order an additional dust filter. See the ISQ and TSQ GC-MS
Spare Parts Guide for ordering information.
1. Shut down your system by following the steps in
on
page
127.
Note You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T10 Torxhead screwdriver to remove the six screws holding the front wire mesh
filter to the interior front panel.
3. Remove and replace the front wire mesh filter.
4. Reattach the six screws holding the front wire mesh filter to the interior front panel.
Figure 417. Replacing the Front Wire Mesh Filter
5. Restart your system by following the instructions in
"Shutting Down the TSQ 9000 System"
Screws
"Restarting the System"
on
page
145.