Thermo Scientific TSQ 9000 Hardware Manual page 248

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Ion Source
220
TSQ 9000 Hardware Manual
.
Note Do not push down on the handle of the source exchange tool while inserting the
ion source cartridge. This could result in opening an air leak that may cause your
instrument to shut itself down to protect the vacuum pumps.
CAUTION INSTRUMENT DAMAGE: Make sure the vacuum interlock handle is all the way
up. Otherwise if you insert the tool with the handle down, you can damage the
instrument.
14. Twist the handle of the source exchange tool to the right and pull it back toward you and
down the right side of the barrel's metal track until the line at the end of the track appears.
Figure 214. Pulling the Tool Down the Right Track
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents