Removing The Source Holder - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
64
TSQ 9000 Hardware Manual
5. Gently pull the source holder away from the ion source cartridge.
Figure 57. Removing the Source Holder
Tip When the source exchange tool is inverted, the source should stay attached. If this
does not happen, the source will not insert correctly into the instrument.
6. Turn the source exchange tool around so that the ion source cartridge is furthest away
from you.
7. Firmly grasp the black handle of the ion cartridge tool in one hand and use your other
hand to pull the metal barrel out and away from the handle so that tool is fully
extended.The ion source cartridge is now hidden inside the barrel.
8. When the barrel can go no further, twist the handle to the left to lock it into position.
9. Insert the barrel end of the ion cartridge tool into the vacuum interlock and twist it to the
right to lock it into position. Be sure the black handle remains in the locked position.
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents