Advanced Troubleshooting
Replacing Fans and Filters
404
TSQ 9000 Hardware Manual
9. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately two inches from the top of the board directly under the white label.
Figure 430. Removing the Lens Driver Board
10. Loosening the screw will separate the board from the instrument. The screw is captured,
so do not attempt to completely remove it from the board and bracket.
11. Lift the lens driver board out of the instrument. Be careful as space is limited.
CAUTION - INSTRUMENT DAMAGE Limit handling of the boards as much as possible.
Do not touch any board components. Hold the board by the sheet metal bracket only.
Lens Driver Board Screw