3
Advanced Troubleshooting
Replacing Components of the Ion Source
Lens Plate and Springs
190
TSQ 9000 Hardware Manual
To replace the lens plate and springs
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
127.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 9000 Instrument"
3. Use a T10 Torxhead screwdriver to disconnect the lens plate and springs from the lens
heater block.
Figure 173. Disconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Plate and
Springs
Screw
4. Disconnect the lens 1, lens 2, and lens 3 wires from the source interface board using a
flat-tipped screwdriver.
"Shutting Down the TSQ 9000 System"
on
page
129.
Thermo Scientific