Replacing The Q3 Quad Assembly Wires - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
Figure 138. Replacing the Q3 Quad Assembly Wires
Screws
5. Attach new wires to each side of the rods. Ensure that the copper washers are on either
side of the wires.
CAUTION - INSTRUMENT DAMAGE Be careful when removing and inserting the screws
on the metal quadrupole rod endcaps. The screws are susceptible to falling between the
rods.
ATTENTION DOMMAGES À L'INSTRUMENT Soyez prudent lorsque vous enlevez et insérez
les vis au niveau des bouchons de tige du quadripôle métallique. Ces vis peuvent tomber
entre les barres.
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Screws
TSQ 9000 Hardware Manual
165