Thermo Scientific TSQ 9000 Hardware Manual page 301

Table of Contents

Advertisement

Thermo Scientific
3. Open the manifold door and replace the magnet yoke, source retainer magnets, or
manifold o-ring as needed on the back of the door,
Figure 295. Replacing the Magnet Yoke, Manifold O-ring, and Source Magnets
Manifold O-ring
Magnet Yoke
4. Close the manifold door and reattach the four screws.
5. Restart your system by following the instructions in
Replacing Components of the Manifold
Source Retainer Magnets
"Restarting the System"
TSQ 9000 Hardware Manual
3
Advanced Troubleshooting
on
page
145.
273