Finding The Lens Wires Between The Quadrupoles; Removing The Q3 Quad Assembly From The Analyzer Tray - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
Figure 102. Finding the Lens Wires between the Quadrupoles
CAUTION INSTRUMENT DAMAGE Quadrupole surface quality is critical for instrument
performance. Take extreme care in handling the quadrupole components.
ATTENTION DOMMAGES À L'INSTRUMENT La qualité de la surface du quadripôle est
essentielle pour les performances de l'instrument. Soyez extrêmement prudent lorsque vous
manipulez les composants de quadripôle.
2. Remove the screws on the Q3 exit endcap holding it to the tray.
3. Remove the Q3 quad assembly and Q3 exit lens assembly from the tray.
4. Detach the Q3 exit lens assembly from the Q3 quad assembly.
5. Set the Q3 quad assembly and Q3 exit lens assembly on a clean work surface.
Figure 103. Removing the Q3 Quad Assembly from the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
6. Remove the three screws holding the collision cell to the tray.
7. Remove the screw connecting the tray to the collision cell grounding wire. See
Figure
104.
8. Remove the collision cell assembly and Q1 quad assembly from the analyzer tray.
3
Replacing Components of the Analyzer
Lens Wires
TSQ 9000 Hardware Manual
Advanced Troubleshooting
Q3 Exit Lens
Assembly
Screws
137