13. Pull the vacuum interlock handle down to close the interior ball valve.
Figure 40. Pulling Down the Vacuum Interlock Handle
14. Twist the end of the vacuum interlock handle and flip the swing handle up so that it rests
next to the vacuum interlock knob location. Then twist the end again to hold it in place.
15. Replace the plug in the VPI to prevent accidental venting and excessive leaks.
CAUTION INSTRUMENT DAMAGE: Do not push the handle all the way up or you will
open the interior ball valve, which will cause you to lose vacuum and possibly damage the
instrument.
ATTENTION DOMMAGES À L'INSTRUMENT : ne relevez pas totalement la poignée ou vous
ouvrirez le robinet à tournant sphérique intérieur, ce qui entraînera une perte de vide et de
possibles dommages à l'instrument.
16. Remove the tool from the instrument.
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
TSQ 9000 Hardware Manual
51