Advanced Troubleshooting
Replacing Components of the Analyzer

Removing the Analyzer Tray Components

136
TSQ 9000 Hardware Manual
24. Move the large tab on the analyzer tray to the left, moving the ion source away from the
transfer line, and then carefully pull the tray toward you through the manifold door.
CAUTION - INSTRUMENT DAMAGE Do not scratch any of the sealing surfaces. Be
especially careful when the source interface board reaches the o-ring surface.
ATTENTION DOMMAGES À L'INSTRUMENT Ne rayez pas les surfaces d'étanchéité. Soyez
particulièrement prudent lorsque la carte d'interface de la source entre en contact avec la
surface du joint torique.
Figure 101. Sliding the Analyzer Tray Out of the Instrument
25. Set the tray on a clean work surface.

To remove the analyzer tray components from the tray

1. Turn the analyzer tray to access the bottom and disconnect the three lens wires attached
to the lenses between the quadrupoles.
Sealing Surfaces
Analyzer Tray
Tab