Thermo Scientific TSQ 9000 Hardware Manual page 407

Table of Contents

Advertisement

Thermo Scientific
Figure 403. Opening the Manifold Door
7. Open the manifold door.
8. Disconnect the calibration and reagent gas tubes from the gas mixing chamber.
9. Remove the screw on the lower right side that is holding the analyzer tray to the manifold.
Figure 404. Disconnecting the Analyzer Tray from the Instrument
3
Advanced Troubleshooting
Replacing the Calibration Gas Components
Screws
TSQ 9000 Hardware Manual
Calibration and
Reagent Gas
Tubes
Right
Screw
Hole
379