Magnet Yoke, Manifold O-Ring, And Source Retainer Magnets - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Manifold

Magnet Yoke, Manifold O-Ring, and Source Retainer Magnets

272
TSQ 9000 Hardware Manual
Note If you have an Advanced EI source, follow the steps in
page 420
to replace these parts.
To replace the magnet yoke, manifold o-ring, or source retainer magnets on the
manifold door
Note For information about replacing the tray alignment pin on the back manifold plate,
see
Back Manifold
Plate.
1. Shut down your system by following the steps in
on
page
127. You do not have to remove the ion source cartridge or the column. If you
are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 294. Opening the Manifold Door
"Advanced EI Ion Source"
"Shutting Down the TSQ 9000 System"
Screws
on
Thermo Scientific