Thermo Scientific TSQ 9000 Hardware Manual page 196

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Analyzer
168
TSQ 9000 Hardware Manual
Figure 141. Replacing the Collision Cell and Quadrupole 3 in the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
7. Reattach the wire to the Q3 exit lens.
8. Replace the analyzer tray into the instrument by following the instructions in
the Analyzer Tray in the TSQ 9000 Instrument"
9. Restart your system by following the instructions in
Q3 Exit Lens
Assembly
on
page
142.
"Restarting the System"
Screws
"Replacing
on
page
145.
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents