Thermo Scientific
7. Move the TSQ 9000 instrument away from the GC so you can access the transfer line on
the right side.
WARNING BURN HAZARD: The transfer line may be hot, so be careful.
AVERTISSEMENT RISQUE DE BRÛLURE : la ligne de transfert peut être chaude, soyez donc
prudent.
8. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
Figure 408. Removing the Right Side Panel Screws
9. Slide the right side panel to the right of the instrument and lift it off.
10. Remove the rear RF board. The rear RF board is the third board standing on end toward
the center of the TSQ 9000 instrument. Use a T20 Torxhead screwdriver to remove the
screw holding the board and bracket to the chassis.
CAUTION - INSTRUMENT DAMAGE Limit handling of the boards as much as possible.
Do not touch any board components. Hold the board by the sheet metal bracket only.
ATTENTION DOMMAGES À L'INSTRUMENT Limitez au maximum la manipulation des
cartes. Ne touchez aucun composant de carte. Gardez la carte uniquement près du support
en tôle.